鉗道管漏水:是由于使用活檢鉗等附件不當(dāng)造成鉗道破損,屬 C 級修理 。
Leakage of forceps: It is caused by improper use of biopsy forceps and other accessories, resulting in damage to the forceps and classified as a C-level repair.
導(dǎo)管束斷:使用年限較長,彎曲管不良造成,屬 C+ 導(dǎo)光束級修理 。
Tube bundle breakage: caused by long service life and poor bending of the tube, it belongs to C+beam level repair.
彎曲角度不到位:鋼絲前端斷:使用年限長,磨損造成, C 級修理 。
Inadequate bending angle: Broken front end of steel wire: Long service life, caused by wear and tear, C-level repair.
彎曲角度不到位:鋼絲后端脫焊和使用一定時間后鋼絲變長,D 級修理,
Inadequate bending angle: The rear end of the steel wire is detached and the steel wire becomes longer after a certain period of use, D-level repair,
(在彎曲好的情況)。
(In a well bent condition).
彎曲橡皮漏水:碰傷造成比較多, D 級修理更換專用橡皮即可 。
Bending rubber leakage: It is caused by more scratches. D-level repair can replace the specialized rubber.
噴嘴堵塞:清洗不干凈, D 級修理更換噴嘴 。
Nozzle blockage: cleaning not clean, D-level repair to replace nozzle.
導(dǎo)像束黑點多或龜裂:纖維鏡使用年限長或保養(yǎng)不當(dāng)造成,A 級修理 。
Multiple black spots or cracks in the imaging beam: caused by long service life or improper maintenance of the fiber mirror, A-level repair.
電子內(nèi)窺鏡鏡頭模糊:具體故障需經(jīng)漏水測試,烘干,拆鏡檢查仔細(xì)判別才能定如何修理(嚴(yán)重時要做 A 級修理)。
The lens of the electronic endoscope is blurry: specific faults need to be carefully judged through water leakage testing, drying, and disassembly inspection before determining how to repair (in severe cases, A-level repair should be performed).
電子鏡打角度時有網(wǎng)狀出現(xiàn):使用年限長 CCD 根部導(dǎo)線斷造成,那一定要做 A 級大修。
When adjusting the angle of the electronic mirror, there is a mesh like appearance: it is caused by a broken wire at the root of the CCD due to its long service life. Therefore, it is necessary to undergo an A-level major repair.